Model: DIM-572
A new generation of modern digital multifunction testers is born. This tester has a range of new features not even found in expensive advanced test equipments. Not only rugged, but designed to excel in harsh environment, still, remaining low cost and affordable. It can be operated with rechargeable batteries, alkaline or low cost general purpose batteries. This unique product has features ranging from Insulation Resistance Testing, Voltage(ac-dc) measurements with Automatic Hold facility, Continuity Test with a short circuit current of minimum 210mA. Two very unique features are found on this multifunction model DIM-572.
MOV and Gas Arrester Testing
Today, most equipments and electrical installations are protected by MOVS and
Gas Arresters, or a combination of these.
This tester can test the devices to establish if it is still operating correctly or not.
Energy conservation is featured on all these new advanced product. “ Auto-power-off” limits the test duration to about 10 seconds to save energy. “Auto-power-off” can be turned off by simply keeping the ON/TEST key depressed for more than 3 seconds while starting a test. This new tester has no moving parts. All calibration are saved internally in a non volatile memory. Calibration can be done at any calibration facility in the world without the need for dedicated calibration equipment. This makes tester easier to maintain and lower the cost of calibration and maintenance. The calibration interval can be extended without much problem. They comply to all the latest regulations, including UK.This tester feature heavy duty protections in its circuitry. Safety features are built-in. For example, should you connect the leads onto a Voltage, AC or DC, the tester will detect it. Switch itself ON and show the voltage present on its leads. The tester will also buzz to inform the user of a safety hazard. We bring a similar solution in the case where the user tries to start a test and measure on a live circuit. If this happen, the tester will disable testing until the circuit has been isolated. This tester can function with rechargeable or Ni-Cad or normal batteries.The circuitry can acept a wide range of power and still perform properly.
Insulation |
|
Test Voltage |
250V/ 500V/ 1000V |
Output Voltage @1mA |
Rated test Voltage +10% Max |
Measuring Range |
2K-2G/ 4K-4G/8K-8G Ohm |
Accuracy |
> 1M ohm +/-3% reading30K Ohm ~ 1M ohm +/-5% reading< 30 K ohm +/-2.0% reading |
When Voltage is constant, current is limited at |
|
Short Circuit current |
+/- 4mA Max |
Continuity Test |
|
Measuring ranges |
0.01 ~ 1999 Ohm |
Accuracy |
0.02 ~ 100 Ohm : +/- 1.0% reading100~300 Ohm : +/- 1.5% reading300 ~ 1999 Ohm : +/- 2.0% reading |
Short Circuit Current |
>220mA |
Open Circuit Voltage |
5V DC |
Resolution |
+/- 2 count |
Voltmeter |
|
AC Voltage (Auto) |
0 ~ 700 V |
DC Voltage (Auto) |
0 ~ 950 V |
Accuracy |
+/- 1.5% of Full Scale |
Resolution |
+/- 1V
|
MOV Test |
|
Test Voltage |
5 ~ 1020V dc |
Voltage Result Accuracy |
+/- 3% |
Voltage Result Resolution |
+/- 2 counts |
GAS Arrester Test |
|
Test Voltage |
5 ~ 1020V dc |
Voltage Result Accuracy |
+/- 3% |
Voltage Result Resolution |
+/- 2 counts |
Protections |
|
Overload |
700V ( Between all terminals) |
Over Voltage |
Class III- 700V toward ground |
Fuses |
1 x 500 mA, 5 x 20mm, HBC, 600 Fas blow |
General |
|
Display |
2 lines x 16characters LCD |
Auto-Null threshold |
5 Ohm |
Buzzer threshold |
3 Ohm |
Fast Test |
10 sec. |
Long test |
60 sec. |
With PI and DAR test function |
10 Min |
Load Battery Test current |
about 300mA |
Power Source |
1.5 x 8 ( Alkaline) |
Dimensions |
175(L) x 85 (W) x 75 (H) mm |
Weight |
655 g Approx. (battery included) |
Operating temperature |
1 deg C to 55 deg C no in full sun |
Storage Tempeatre |
-20 deg C to + 70 deg C |
Accessories |
Test leads( TL-AL11-A), Shoulder belt, instruciton manual, Batteries |