TECPEL - Propel Quality Measurement.

 

TECPEL-Test and Measurement

  Tecpel Product Guide
 

 

 3MHz  Function Generator + 60MHz Freq. Counter

Model : FG-32

 

  

   FEATURES:

6 function, 6 ranges
 Frequency Range 0.5Hz~3MHz
Linear / Log sweep fucntion.
 Frequency  Counter  0.2 ~ 60MHz
 Frequency Counter Max. Resoltuion 0.001Hz?C

  

  Leaflet DM-1        Data Sheet DM-2  

               

               

                

FG 32 (6  waveforms , 6 ranges ,  Freq. Counter)  :  specification   CE

 Main Output

?@Frequency Range

0.5Hz ~ 3MHz in 6 ranges , 5 digits LED display, High resolution 0.001Hz  ?Ause rotary switch to select ranges
?@Waveform 6 Waveforms ( Sins, Square, Triangle, Ramp, + Pulse, - Pulse), Use rotary switch to select waveform
?@Stability 0.1% after 15 minutes;?@?@0.2% after 24 hours
  DC OFFSET    +/- 10 V with pull Adjust.  switch  ;  + - 10V( No load)?A + - 5V(50?[ Load)
 Ramp
   Frequency Range 0.5Hz ~ 2.5MHz?A Use 5 digits LED display
  Max. Resolution 0.001Hz
?@Symmetry 50%(Rise waveform)  to 50%( Fall waveform) Distortion??1%(1Hz ~ 100KHz)
?@Distortion ??1%(1Hz ~ 100KHz)
 Triangle
?@Frequency Range 0.5Hz ~ 3MHz?A   Use 5 digits  LED display
?@Max. Resolution 0.001Hz
    Symmetry 50%(Rise waveform)  to 50%( Fall waveform) Distortion??1%(1Hz ~ 100KHz)
?@Distortion ??1%(1Hz ~ 100KHz)
 Sine Ware
?@Frequency Range 0.5Hz ~ 3MHz?AUse 5 digits  LED display
?@Max. Resolution 0.001Hz
?@ Symmetry 50%(Rise waveform)  to 50%( Fall waveform) Distortion??1%(1Hz ~ 100KHz)
?@Distortion ??1%  to100KHz
?@Harmonic Ratio ??30dB  to3MHz
?@ Freq.  Response ??0.1dB  to 100MHz?@  ??1dB  to 3MHz
 Square
?@Frequency Range  0.5Hz ~ 3MHz?AUse 5 digits  LED display
?@Max. Resolution 0.001Hz
?@ Symmetry 50%(Positive half circle) to 50%( Negative half cycle) distortion ??1%  to 100KHz
?@Rise Time ??60nS
  + Pulse
?@ Freq. Range 0.5Hz ~ 3MHz?AUse 5 digits  LED display
?@Max. Resolution 0.001Hz
?@+ Pulse width 0.4sec ~ 100nS(Continously variable)
?@Symmetry 50%(Positive half circle) to 50%( Negative half cycle) distortion ??1%  to 100KHz
?@Rise Time ??60nS
  - Pulse
?@Freq. Range  0.5Hz ~ 3MHz?AUse 5 digits  LED display
?@Max Resolution 0.001Hz
?@- pulse width 0.4sec~100nS
?@Symmetry 85%(+ time base) to 15%(- Pulse) distortion ??5% to 100KHz
?@Fall time ??60nS
  Main Output
?@Output Impedance 50?[?A Distortion ??2%
?@Max. Output 20Vp-p( No load)?A0.05Vp-p(50?[ Load)?@?@Distortion?G + - 0.5V
?@Min. Output 0.1Vp-p( No load)?A0.05Vp-p(50?[ load)
?@Attenuator -20dB  Attenuator ; Attenuation ??2%
 SYNC OUTPUT
?@Output Impedance 50?[?ADistortion ??2%
?@Output Level TTL Level ??3Vp-p Fix Amplitude
?@Rise Time ??40nS
VCF  Input
?@Input Level 0~10V Input
?@Max. Freq. Variable Rate 1 : 1 ~ 1 : 1000
    Input Frequency DC ~ 1KHz
  Sweep Sync Output
?@Output Impedance 1K?[?ADistortion ??2%
?@Output Waveform  Linearity or LOG sweep  or LOG sweep slope
?@Output Amplitude 10Vp-p(load)?A5Vp-p(51K?[  Load)
?@Output Frequency   Continuously variable   , 0.2Hz ~ 100Hz
Sweep
?@Sweep Mode LINEAR/ LOG  sweep,  Use switch selector
?@Sweep Rate 5sec ~ 10mS?AContinuously variable
    Sweep Width 1 : 1 ~ 1 : 100( To cooperate with Freq. Knob)
  Frequency Counter
?@Display 5  digits 0.36 '   Red  LED display
?@Max. Resolution 0.001Hz
?@Unit Hz/KMz
?@ TIME BASE 20MHz
?@ Temp. Characteristic ??10PPM
?@Distortion ??0.002% ?? ?@??
?@Power source +5V / 160mA
  Internal Counter
?@ Range  Auto Range (0.001Hz/0.01Hz/0.001KHz/0.01KHz)?@
 Total 4 ranges , Resolution ?ACPU select one to display
?@Counter Range 0.500Hz ~ 3000.0KHz?AAuto select decimal point and display unit
?@Gate Time Variable  0.25sec ~ 2sec  Auto set
?@ Minimum Display  4 digits
   External Counter
   Max. Input Voltage ??250Vrms
?@Input Impedance 1M  ,  Distortion ??2%
?@Input Frequency Range 0.2Hz ~ 60MHz
?@ Attenuation switch x20(-26nB) 
   (COUPLING) AC(HF)-- Measure  frequency over 100KHz through capacitor coupling
DC(LF)100KHz FILTER-- Measure frequency under 100KHz direct coupling
    (GATE TIME)  Variable 0.25sec ~ 10sec auto set by frequency
    Range  Auto range(0.001Hz/0.01Hz/0.001KHz/0.1KHz/1KHz)  5 kinds of resolution  selected by CPU.
?@Minimum display  4 digits
?@Input sensitivity ??30mVrms(1MHz)

Calculation  Time table for Frequency Display value/ Gate time

Standard Value GATE TIME INT EXT.HF EXT.LF
0.2Hz 10S ?? ?? 0.200Hz
0.5Hz 2S 0.500Hz ?? 0.500Hz
1Hz 1S 1.000Hz ?? 1.000Hz
10Hz 0.25S 10.000Hz ?? 10.000Hz
100Hz 0.25S 100.00Hz ?? 100.00Hz
1KHz 0.25S 1.000KHz ?? 1.000KHz
10KHz 0.25S 10.000KHz ?? 10.000KHz
100KHz 0.25S 100.0KHz 100.00KHz 100.0KHz
1MHz 0.25S 1000.0KHz 1000KHz ??
10MHz 0.25S ?? 10000KHz ??
60MHz 0.25S ?? 60000KHz ??
 General Specifications:
?@Operating Temp.  Temperature: 0 ~ 40 C  ;   Humidity 0 ~ 90% RH
?@Power Source AC 115V(+-10%) 50/60Hz , FUSE 600mA?C
AC 230V?@(+-10%) 50/60Hz , FUSE 300mA
Consumption Max 25VA(Watt)
?@Heat Ventilation  Use fan  DC 12V/100mA
?@Safety Approval CE,GS,UL,CUL  approval
?@Dimenstions 275mm(Width) x 90mm( Height) x 300mm(Depth)
    Weight 2.5Kg( Net)
?@Accessories  Operation Manual, Test leads, Power cord.

?@

 LEAFLET DM # 1

?@

 Data Sheet  DM  # 2

 


 

 

 

 

 

 

TECPEL CO., LTD.

E-MAIL: sales@tecpel.com ; oka@oka.com.tw
Tel : 886 (2)2737-5866 FAX : (02)2737-3343
Web-site: http://www.tecpel.com; http://www.oka.com.tw

 