General Specifications:
- Circuit Custom one-chip of microprocessor LSI PH Optional,
- circuit. Electrode Any PH electrode with BNC connector.
- Display 51 mm x 32 mm, dual function LCD display,
- 15 mm ( 0.6" ) digit size. Operating 0 to 50 (32 to 122 )°F
- Measurement PH 0 to 14 PH Temperature
- mV -1999 mV to 1999 mV Operating Max. 80% RH.
- Input 12 Humidity
- Impedance 10 ohm Power Supply 006P DC 9V battery (Optional)
- Temperature Manual 0 to 100 °C, be adjusted by ( Alkaline or Heavy duty type ).
- Compensation push button on front panel. Power Current Approx. DC 7 mA.
- for pH Automatic With the optional TEMP. Weight 250 g/0.55 LB ( battery included ).
- measurement ( ATC ) probe ( TP-07 ) Size ( meter ) 195 x 68 x 30 mm ( 7.6 x 2.6 x 1.2 inch ). 0 to 65 ° C.
- Standard Instruction manual......................1 PC.
- pH PH7, PH4, and PH10, 3 points calibration Accessories
- Calibration ensure the best linearity and accuracy. Optional PH electrodes :
- Data hold Hold the reading value on the display. Probes & * General purpose PH electrode, PE-03
- Memory Maximum and Minimum reading values can Accessories * General purpose PH electrode, PE-11
- Recall be saved and retrieved by record function. * Professional PH electrode, PE-01
- Sampling Time Approx. 0.8 second. * Spear tip PH electrode, PE-04HD
- Power off Auto shut off saves battery life, or manual ATC temp. probe, TP-07
- off by push button. PH electrodes + Temp. Probe, PE-03K7
- Data Output RS 232 computer serial interface. ORP electrode, ORP-04
- Overload "- - - -" symbol on the display. Hard carrying case, CA-06
- indication RS232 cable, UPCB-02
- Application Software, SW-U801-WIN
- ELECTRICAL SPECIFICATIONS ( 23 ± 5 )°C
- ELECTRICAL SPECIFICATIONS ( 23 ± 5 ) °C
- pH meter specifications
- Measurement Measuring Range Resolution Accuracy
- PH 0 to 14 PH 0.01 PH ± (0.02 PH + 2 d) * Meter only
- mV -1999 mV to 1999 mV 1 mV ± (0.5% + 2 d) * Meter only
- Temperature Compensation for pH measurement :
- * Manual : 0 to 100 , be adjusted by push button on front panel.
- * Automatic ( ATC ) : With the optional TEMP. probe ( TP-07 ) 0 to 65 °C
- pH Calibration :
- * PH7, PH4, and PH10, 3 points calibration ensure the best linearity and accuracy.
- APPLICATIONS
- APPLICATIONS
- * AQUARIUMS * FOOD PROCESSING * PAPER INDUSTRY * SCHOOLS & COLLEGES
- * BEVERAGE * PHOTOGRAPHY * PLATING INDUSTRY * SWIMMING POOLS
- * FISH HATCHERIES * LABORATORY * QUALITY CONTROL * WATER CONDITIONING
|